Call us

Product documentation/UHC Heater Chiller -40°C to 200°C 10/40 20/40 50/40

UHC Heater Chiller -40°C to 200°C 10/40 20/40 50/40

UHC Heater Chiller -40°C to 200°C 10/40 20/40 50/40

Documents for UHC Heater Chiller -40°C to 200°C 10/40 20/40 50/40, covering models UHC, UHC-2040, UHC-1040 and UHC-5040: a user manual, 4 spec sheets and a compliance document. Last updated Oct 23, 2026.

Documents 6

About this product

The UHC Series uses both heating and cooling to provide a regulated temperature fluid output from -40°C to 200°C with +/- 2°C accuracy, for use with reactor jackets or other heat exchangers. A plate heat exchanger achieves large changes in temperature rapidly, and the closed loop reservoir, casters, and included insulated metal flex hoses support practical lab setups.

  • -40 to 200 °CTemperature range
  • 6,247 to 33,058 BTUCooling at 20°C
  • 3 to 6 kWHeating wattage
  • 18 to 35 L/minPump flow
  • 6 to 12 LReservoir capacity
  • Heating and cooling. Both work together for a regulated fluid output, +/- 2°C accuracy.
  • Rapid temperature changes. Plate heat exchanger achieves large changes in temperature rapidly.
  • Closed loop reservoir. Small in-feed hole closes the reservoir off from the outside environment.
  • Insulated metal hoses. Two 3/4" insulated 6ft hoses included for higher and lower temperatures.
  • Mobile on casters. Casters aid portability, with easily accessible control panels.

Applications: Jacketed reactors, Rotary evaporators, Vacuum systems, Columns, Solvent tanks, Analytical instruments

Specifications

Temperature Control
Temperature range-40°C to 200°C
Temperature accuracy+/- 2°C
Control methodHeating and cooling for regulated fluid output
Recommended fluidsSilicone oil (-40°C to 200°C) or PG 60% : DW 40% (-40°C to 110°C)
Refrigeration
RefrigerantR-448 (2025+)
CompressorPhase change compressor
Heat exchangerPlate heat exchanger
Refrigerating wattage1.9kW to 10.2kW
BTU capacity @20°C6,247 to 33,058
BTU capacity @-40°C3,344 to 8,737
Operation
ControlsLCD control panel on front
Duty cycleMaximum operation time 16 hours; not for 24-hour operation
Operating ambient70°F or 20°C
Clearance12-16 inches around the unit
MobilityCasters
Circulation
Pump flow18L/Min to 35L/Min
Pump pressure6.5 PSI to 28.48 PSI
Pump wattage100W to 250W
Fluid connection3/4" MNPT
Loop typeClosed loop
Electrical
Power supply220-240V 60Hz Single Phase
Current30A to 45A
Wattage5kW to 10kW
Heating wattage3kW to 6kW
ConnectionNEMA 6-20P, NEMA 6-30P, or hardwired with quick disconnect panel
Voltage range219V - 253V; not compatible with 208V
Bath and Dimensions
Reservoir capacity6L to 12L
Dimensions25" x 22" x 42" to 28" x 26" x 44"
Weight412 lb
Included hoses3/4" Insulated Tubing - 6ft, 2 pcs

From the approved spec sheet. Download the spec sheet PDF.

Request a document

Tell us which document and model number you need and we'll send it over, usually the same business day.

Loading the form…