The UHC Series uses both heating and cooling to provide a regulated temperature fluid output from -40°C to 200°C with +/- 2°C accuracy, for use with reactor jackets or other heat exchangers. A plate heat exchanger achieves large changes in temperature rapidly, and the closed loop reservoir, casters, and included insulated metal flex hoses support practical lab setups.
- -40 to 200 °CTemperature range
- 6,247 to 33,058 BTUCooling at 20°C
- 3 to 6 kWHeating wattage
- 18 to 35 L/minPump flow
- 6 to 12 LReservoir capacity
- Heating and cooling. Both work together for a regulated fluid output, +/- 2°C accuracy.
- Rapid temperature changes. Plate heat exchanger achieves large changes in temperature rapidly.
- Closed loop reservoir. Small in-feed hole closes the reservoir off from the outside environment.
- Insulated metal hoses. Two 3/4" insulated 6ft hoses included for higher and lower temperatures.
- Mobile on casters. Casters aid portability, with easily accessible control panels.
Applications: Jacketed reactors, Rotary evaporators, Vacuum systems, Columns, Solvent tanks, Analytical instruments
Specifications
| Temperature Control | |
|---|---|
| Temperature range | -40°C to 200°C |
| Temperature accuracy | +/- 2°C |
| Control method | Heating and cooling for regulated fluid output |
| Recommended fluids | Silicone oil (-40°C to 200°C) or PG 60% : DW 40% (-40°C to 110°C) |
| Refrigeration | |
| Refrigerant | R-448 (2025+) |
| Compressor | Phase change compressor |
| Heat exchanger | Plate heat exchanger |
| Refrigerating wattage | 1.9kW to 10.2kW |
| BTU capacity @20°C | 6,247 to 33,058 |
| BTU capacity @-40°C | 3,344 to 8,737 |
| Operation | |
| Controls | LCD control panel on front |
| Duty cycle | Maximum operation time 16 hours; not for 24-hour operation |
| Operating ambient | 70°F or 20°C |
| Clearance | 12-16 inches around the unit |
| Mobility | Casters |
| Circulation | |
| Pump flow | 18L/Min to 35L/Min |
| Pump pressure | 6.5 PSI to 28.48 PSI |
| Pump wattage | 100W to 250W |
| Fluid connection | 3/4" MNPT |
| Loop type | Closed loop |
| Electrical | |
| Power supply | 220-240V 60Hz Single Phase |
| Current | 30A to 45A |
| Wattage | 5kW to 10kW |
| Heating wattage | 3kW to 6kW |
| Connection | NEMA 6-20P, NEMA 6-30P, or hardwired with quick disconnect panel |
| Voltage range | 219V - 253V; not compatible with 208V |
| Bath and Dimensions | |
| Reservoir capacity | 6L to 12L |
| Dimensions | 25" x 22" x 42" to 28" x 26" x 44" |
| Weight | 412 lb |
| Included hoses | 3/4" Insulated Tubing - 6ft, 2 pcs |
From the approved spec sheet. Download the spec sheet PDF.